Repair 4D Philips V6-2 IU22/IE33 Cable/Connecting Cable Leakage/Damaged Cable
Cable leakage failure reason and solution
Ultrasound probe cable is adopted by high quality of high multi-fiber shielded cable, and good workmanship wires, many hundreds of root, like hair thickness. Due to the bending and torsion of the cable, it makes outside of reserve cable insulating layer will be broken, fracture, revealing the inside line, shielding layer damage, break, corrugated and defect image can be interferenced.
Solution
Due to probe affected by the external force or wrong operators, the probe was not properly placed within for a long time, cable conductor faults. Open the outer leather, lateral horizontal gashing the total shielding wire , pick out the deep crack line, welding is completed, with transparent tape around one layer, use a multimeter to measure whether it is turned on,this wire shielding will be connected and wrapped the tape, to connect the wire shielding and tie up good adhesive tape. As above, If multiple root fracture. After commissioning, everything is all right to better the outermost shielded wire welding, fixed outer holster with high voltage margin plastic tape. in this way, you can continue to use the probe for a long time.