Refurbished OCULUS Easyfield

Refurbished
6
USA

Seller:

USA

Product Details

OCULUS
Easyfield
Ophthalmic perimeters
17.04.2024
248

Product Specifications

10000 asb
Goldmann III
30
Kinetic
0.2
31.4 asb
Show

Product Description

Programs: Pre-defined glaucoma, macula, screening and neurological tests
User-defined tests
Test patterns: 30-2, 24-2, 30x24, 10-2, hemisphere, customized patterns
Strategies: Threshold strategies: SPARK Quick, CLIP, OCULUS Fast Threshold, Full Threshold (4/2)
Optional: SPARK Precision
Age adapted suprathreshold screening (2-zone, 3-zone, quantify defects)
Examination Speed: Adaptive, fast, normal, slow, user-defined
Fixation control: CMOS camera, through central threshold, Heijl-Krakau (using the blind spot)
Result Display: Glaucoma Staging System (GSS2), Glaucoma Staging Program (GSP)
Reports: Threshold Noiseless Trend (TNT) progression report
Specifications Perimeter bowl radius: r = 30 cm (11.8 in)
Maximum eccentricity: 30°
Stimulus size: Goldmann III
Stimulus luminance range / increments: 0.03 – 3,180 cd/m² (0.1 – 10,000 asb) / 0.1 log steps
Background luminance: 10 cd/m² (31.4 asb)
Stimulus color: White
Stimulus duration: 200ms / user-defined
Patient positioning: Measurement head with adjustable angle of inclination,
adaptable double chinrest, double headrest
Software: Device control, patient management, backup, and print software (Windows®)
Built-in networking, easy EMR-integration, DICOM compatibility
Interface: USB, RS232
Technical specification Dimensions (W x D x H) 316 x 506 x 435 mm (12.4 x 19.9 x 17.1 in)
Weight with accessories 6.8 kg (15.0 lbs)
Max. power consumption 30 W
Voltage and
frequency 100/240 V AC
50-60Hz
Min. computer requirements Operating system: Windows® XP