New PEAK SCIENTIFIC Corona 1010

New
1
In stock
Netherlands
6 473
(plus shipping costs and VAT)
Price in seller currency:
6 000

Seller:

Netherlands

Product Details

PEAK SCIENTIFIC
Corona 1010
Nitrogen generator
27.06.2023
429

Product Description

Corona Nitrogen 1010

Our Corona 1010 has been engineered to meet the specifications of the Thermo Scientific Corona Veo Charged Aerosol Detector. This system can be purchased as a single nitrogen unit for laboratories with an external air supply or with a stackable air compressor for laboratories without an external air supply. The Corona system is also suitable for a variety of other applications, requiring up to 5 L/min nitrogen for LC, HPLC or UHPLC.

The Corona 1010 has been developed by Peak Scientific in conjunction with Thermo

Fisher Scientific to supply pure Nitrogen to their Charged Aerosol Detectors (CAD).

The Corona 1010 generates gas at the right flow, pressures and purities required by the

Charge Aerosol Detector and has been fully tested and approved by Thermo.

Peak Scientific offers a compact, stackable solution with their Corona products, but

also with a Corona Veo CAD itself.

Other features include:

• Generates Nitrogen on demand from compressed air.

• Effective and reliable membrane technology.

• Minimum maintenance with only a filter change per annum.

• Small and Stackable.

To ensure these generator models meet our high expectations with regards to reliability

and performance, we have tested them extensively at our manufacturing plant and with

end users around the world to ensure reliability and longevity of the system.

Air Purity

The Corona 1010 generator should be connected to an air supply that, as a minimum,

meets ISO8753-1:2010 Class 1.4.1

Class 1 Particulate

In each cubic metre of compressed air, the particulate count should not exceed 20,000

particles in the 0.1 †0.5 micron size range, 400 particles in the 0.5 †1 micron range and

10 particles in the 1 †5 micron size range.

Class 4 Water

A pressure dew point (PDP) of +3В°C or better is required and no liquid water is allowed.

Class 1 Oil

In each cubic metre of compressed air, not more than 0.01mg of oil is allowed. This is a

total level for liquid oil, oil aerosol and oil vapour.

On Demand Gas

The generator will produce nitrogen on demand. If the application is operating

and requires a gas flow, the system it is connected to will supply this to suit the

requirements of the application. If the application requirement for gas stops, the system

will also stop. If the demand from the application starts again, the system will detect

the demand for gas and will automatically start again to suit the demand.

Specifications:

Min Inlet Pressure:В 110psi/7.6bar

Max Inlet Pressure:В 130psi/9bar

Min Air Inlet Flow:В 30

Min Inlet Air Quality:В ISO 8753-1:2010 Class 1.4.1

Start Up Time:В 5 mins

Power Consumption:В up to 23 VA

Voltage:В 100- 230 V

Frequency:В 50 / 60

Current:В 0.1 A

Heat output:В up to 78 BTU

Max Operating Temp:В 35В°C / 95В°F

Particles:В